5-4
Tables 5-1 and 5-2 show guidelines on how to convert hold times
for the DNA Thermal Cycler or DNA Thermal Cycler 480 to hold
times for the GeneAmp PCR System 9600. The table you use will
depend on whether you are using an
up
ramp (the temperature is
higher than in the previous segment) or a
down
ramp (the
temperature is lower).
If you are using an
up
ramp, you will base the new hold time
on the change in temperature required to reach the next
target temperature. See Table 5-1.
If you are using a
down
ramp, you will base the new hold time
on the starting temperature of the ramp and the change in
temperature required to reach the next target temperature.
See Table 5-2.
For up ramps, determine the change in temperature required to
reach the next target temperature and round this off to the closest
value given below. Then subtract the number of seconds
indicated from the hold time used for the DNA Thermal Cycler.
This will give you the hold time to use for the GeneAmp PCR
System 9600. The typically recommended hold time is 10 to 15
seconds for denaturation.
Table 5-1 After Up Ramps: Converting Hold Times Greater than
One Minute (DNA Thermal Cycler to GeneAmp PCR System 9600)
Temp.
(°C)
Seconds to Subtract from DNA Thermal
Cycler Hold Times (> 1 min.)
10° 38 sec.
15° 45 sec.
20° 49 sec.
30° 54 sec.
40° 55 sec.
50° 57 sec.
60° 57 sec.